TY - JOUR T1 - Disc antibiogram test with graded plate method TT - تست آنتی بیوگرام دیسکی با روش پلیت های مدرج JF - mljgoums JO - mljgoums VL - 3 IS - 2 UR - http://mlj.goums.ac.ir/article-1-16-en.html Y1 - 2009 SP - 0 EP - 0 KW - Keywords: Antibiotic KW - Disc diffusion test KW - Dispensing disc KW - Non-growth haloes KW - Non-growth haloes interference KW - Plate N2 - Abstract Background and objectives: Halo interference around antibiogram disc is one of the problems seen in Agar media. This study aimed at applying graded plate method to reduce this problem. Material and Methods: Forty-eight millimeter plates, before putting disc, were graded and the location of discs was determined. Instead of full disc, we used half-disc and then compared the haloe interference, based on clinical and Laboratory Standard Institute (CLSI), of Klebseila Pneumoniae in the presence of nine different antibiotics. Results: The tests performed on Klebseila Pneumoniae show that five antibiotics such as Enrofloxacin (24mm), Flumequine (12mm), Furazolidone (22mm), ampicillin (22mm) and Florfenicle (22mm) cause non-growth haloes while none of them has interference haloe when we use the news method. Conclusion: According to results, the graded plate not only has negative effect on Antibiogram results but also is careful and decrease the interference produced by previous methods and its efficacy is increased up to %17.64 when we use 84 diametered Plate. Keywords: Antibiotic, Disc diffusion test, Dispensing disc, Non-growth haloes, Non-growth haloes interference, Plate M3 ER -