Mohammad-Hosein Validad , Hava Shafiee , Monireh Mahjoob , Tahereh Rakhshandadi , Hadi Ostadimoghaddam , Abbas-Ali Yekta , Javad Heravian Shandiz , Neda Nakhjavanpour , Sara Farsi ,
Volume 27, Issue 2 (Summer 2025)
Abstract

Background and Objective: Evaluating the agreement among various corneal imaging devices is crucial due to their varying technologies in diagnosing corneal parameters. This study aimed to assess the agreement of Advanced Corneal Explorer (ACE) with Pentacam and Orbscan in myopic individuals with healthy corneas.
Methods: This descriptive-analytical study was conducted on 45 myopic individuals (25 women and 20 men; mean age=30.37±6.13 years) referring to Al-Zahra Ophthalmology Hospital in Zahedan, Iran, during 2023. Anterior segment imaging was performed using three devices: ACE, Pentacam, and Orbscan. Corneal parameters, including topography, tomography, and corneal aberrations, were examined. The intraclass correlation coefficient (ICC) and 95% limits of agreement (LoA) were used to evaluate the agreement of ACE data with the data obtained from the other two devices.
Results: ACE exhibited good agreement with Pentacam for the mean power (ICC=0.97), maximum power (ICC=0.98), and minimum power of the anterior corneal surface (ICC=0.97) (P<0.05). Additionally, ACE exhibited good statistical agreement with Orbscan for the mean power of the anterior corneal surface due to an ICC of 0.98 (P<0.05). The ICC values for central corneal thickness and thinnest corneal thickness with the ACE and Pentacam devices were 0.99 and 0.95, respectively, indicating significant statistical agreement (P<0.05). White-to-white distance, pupil size, and corneal aberrations did not show good agreement among the three devices (ICC<0.75).
Conclusion: Topography values and corneal thickness measured with ACE showed good agreement with Pentacam; therefore, they can be used interchangeably. Other measured variables from ACE did not show good agreement with Pentacam and Orbscan and are not interchangeable.